Greetings! In this article, we'll discuss the Techcross BWTS (ECS) ballast system, its operating logic, and troubleshooting instructions from the Korean manufacturer.
The Electro-Cleen System (ECS) by Techcross is an electrochlorination-type Ballast Water Treatment System (BWTS). It disinfects ballast water during uptake by generating sodium hypochlorite (NaOCl) from seawater and neutralizes residual oxidants during discharge to comply with IMO D-2 regulations.
This article explains:
- The complete operating logic of the system
- Functional sequence in Ballasting and De-ballasting modes
- Control philosophy
- A structured TRO troubleshooting algorithm
System Working Principle
Techcross ECS is a side-stream electrochlorination system.
Core Concept:
- During ballasting → active substance (NaOCl) is generated and injected.
- During de-ballasting → residual oxidants are neutralized.
Main Components
- Ballast Water Filter (typically 40 μm)
- Side-stream Electrolysis Unit
- Power Supply Unit (PSU)
- Hydrogen Ventilation System
- TRO Analyzer
- Neutralization Unit (Na₂S₂O₃ dosing)
- Salinity Sensor
- Flow Meter / Flow Switch
- PLC Control Panel (HMI)
Complete Operating Logic
A. Ballasting Mode (Treatment Mode)
Step 1 – Start Conditions
The system will start only if:
- Ballast pump running signal is present
- Flow rate above minimum threshold
- Salinity > 1 PSU
- No active critical alarms
- Hydrogen ventilation operational
If any interlock fails → electrolysis will not start.
Step 2 – Filtration
All ballast water passes through the automatic filter:
- Organisms > 40 μm removed
- Differential pressure monitored
- Automatic backflush activated when required
Step 3 – Electrolysis (Active Substance Generation)
A portion of seawater is diverted into the electrolysis cell.
Chemical reaction: NaCl + H₂O + electricity → NaOCl + H₂
Produced:
- Sodium hypochlorite (disinfectant)
- Hydrogen gas (vented safely)
Step 4 – Injection and Mixing
Generated NaOCl is injected into the main ballast line. PLC controls:
- DC current
- Output capacity
- TRO setpoint
Typical TRO during ballasting: 6–10 mg/L (model dependent)
Step 5 – TRO Feedback Control Loop
The TRO analyzer measures Total Residual Oxidant.
Control logic:
- If TRO low → increase DC current
- If TRO high → decrease DC current
Closed-loop automatic regulation ensures proper dosage.
Step 6 – Holding Time
After tank filling:
- Minimum 24 hours holding time required
- Allows full inactivation of microorganisms < 10 μm.
B. De-ballasting Mode (Neutralization Mode)
Electrolysis is OFF.
Step 1 – Residual TRO Measurement
Water in ballast tank may contain remaining oxidants.
Regulatory discharge limit: ≤ 0.1 mg/L TRO
Step 2 – Neutralization
Sodium thiosulfate (Na₂S₂O₃) is injected.
Chemical reaction: NaOCl + Na₂S₂O₃ → neutralized compounds
Step 3 – Final TRO Monitoring
TRO is measured at discharge line.
If above limit: dosing rate increases automatically.
5. Control Philosophy Summary
The system will not operate if:
- No flow
- Low salinity
- Hydrogen ventilation failure
- TRO sensor failure
- Electrolyzer overcurrent
- PSU overheating
- PRU (Power Rectifier Unit) overheating
Safety interlocks are integrated at PLC level.
TRO Troubleshooting Algorithm
TRO-related faults are the most common. We divide them into:
- LOW TRO
- HIGH TRO
A. LOW TRO During Ballasting
Step 1 – Check Electrolysis Current
Observe HMI:
- DC Current (A)
- DC Voltage (V)
If current = 0 or very low:
- → Check PSU
- → Check interlocks
- → Check contactors and protection devices
If salinity < 1 PSU: system cannot generate sufficient NaOCl.
Verify:
- Salinity sensor reading
- Manual refractometer measurement
Step 3 – Check Side-Stream Flow
Insufficient flow through electrolyzer:
- Blocked strainer
- Closed valve
- Air lock
- Low pressure
Step 4 – Inspect Electrolysis Cell
Possible issues:
- Scaling
- Fouling
- Worn electrodes
- Internal blockage
Cleaning may be required.
Step 5 – Verify TRO Analyzer
Most frequent cause of LOW TRO alarm.
Check:
- Flow cell contamination
- Air bubbles
- Membrane condition
- Calibration date
Perform manual DPD test.
If manual test normal but sensor low → sensor malfunction.
B. HIGH TRO During Ballasting
Step 1 – Check Flow Rate
- Reduced ballast flow increases concentration.
- Verify actual flow vs design.
Step 2 – Check Feedback Loop
If TRO sensor slow or frozen:
- System continues generating.
- Recalibrate sensor.
Step 3 – Check Mixing Quality
Improper injection point may cause localized high readings.
C. LOW TRO During Discharge
Means insufficient neutralization.
Check:
- Na₂S₂O₃ tank level
- Dosing pump stroke
- Injection line blockage
- Check valves
D. HIGH TRO During Discharge (Critical Situation)
High PSC risk.
Immediate checks:
- Neutralizer tank level
- Dosing pump operation
- TRO sensor calibration
- Manual DPD verification
If manual reading normal → analyzer fault.
7. Practical Field Diagnostic Logic
LOW TRO?
Is DC current present?
- NO → PSU / interlock issue
- YES → Is salinity sufficient?
- YES → Is side-stream flow adequate?
- YES → Is electrolyzer clean?
- YES → Is TRO sensor calibrated?
HIGH TRO?
- Is ballast flow stable?
- Is TRO sensor reliable?
- Is mixing adequate?
- Is neutralization working?
8. Most Common Real-Life Causes
- Dirty TRO analyzer
- Low salinity water
- Incorrect calibration
- Side-stream blockage
- Electrolyzer scaling
Techcross ECS BWTS is a robust electrochlorination system built around:
- Side-stream active substance generation
- Closed-loop TRO control
- Automatic neutralization during discharge
Understanding the control logic and TRO feedback mechanism allows engineers to diagnose 90% of operational problems efficiently.
Thank you for your attention! I hope this article was useful.
Manuals! As always, I'd like to draw your attention to the manufacturer's official instructions. This article was written based on the instructions and my personal experience. If you need manuals for the BWTS Techcross, I recommend our closed Telegram channel "Marine Engineering Manuals." In this channel, you'll find not only all the necessary manuals and troubleshooting service manuals, but also video tutorials on how to operate the BWTS Techcross.
If you encounter any issues with this ballast system, please let me know in the comments, and I'll try to help. By the way, there's a helpful article on our website: "ERMA First BWTS. Simple Instructions for Using a Ballast System". It also describes a similar BWMS system with electrolysis.
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)

.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
No comments:
Post a Comment